User Contributed MET/CAL Procedure ============================================================================= INSTRUMENT: 5700A Zero Cal and Cal Check Report to a File DATE: 21-Sep-92 AUTHOR: User Contributed REVISION: 1.0 ADJUSTMENT THRESHOLD: 70% NUMBER OF TESTS: 1 NUMBER OF LINES: 35 ============================================================================= STEP FSC RANGE NOMINAL TOLERANCE MOD1 MOD2 3 4 CON 1.001 DISP This procedure will perform a 3 minute Zero Calibration 1.001 DISP 1 hour Calibration Check and save the Report to a File. 1.001 DISP [32] 1.001 DISP Be sure that the 5700A Calibrator has been in a stable 1.001 DISP environment and turned ON for at least one hour. 1.002 IEEE [@15]*CLS;*RST;*WAI;*OPC 1.003 HEAD [27][91]1m[27][91]5mCAL_ZERO IN PROGRESS![27][91]0m 1.004 IEEE CAL_ZERO;*CLS;*OPC 1.005 IEEE *ESR?[D1000][I] 1.006 MATH MEM1=MEM-1 1.007 JMPZ 1.009 1.008 JMP 1.005 1.009 HEAD [27][91]1m[27][91]5mCAL_CHK IN PROGRESS![27][91]0m 1.010 IEEE CAL_CHK;*CLS;*OPC 1.011 IEEE *ESR?[D1000][I] 1.012 MATH MEM1=MEM-1 1.013 JMPZ 1.015 1.014 JMP 1.011 1.015 HEAD [27][91]5mSAVING CAL_CHK TO FILE CAL_RPT.CHK[27][91]m 1.016 IEEE CAL_RPT? CHECK[I>CAL_RPT.CHK] 1.017 HEAD 1.018 DISP Zero Calibration and Cal Check Completed 1.018 DISP Report saved in file CAL_RPT.CHK 1.019 END